A comparison of the shock-absorbing properties of cervical disc prosthesis bearing materials

Michael C. Dahl, PhD,1 Stephen Jacobsen, MD,2 Newton Metcalf Jr,3 Rick Sasso, MD,4 Randal P. Ching, PhD5
1Disc Dynamics Inc., Eden Prairie, MN 2Department of Orthopedics, Indiana University School of Medicine, Indianapolis, IN 3Medtronic Sofamor Danek, Memphis, TN 4Indiana Spine Group, Indianapolis, IN 5Applied Biomechanics Laboratory, University of Washington, Seattle, WA

Abstract 

Background Data

Cervical arthroplasty offers theoretical advantages over traditional spinal fusion, including elimination of adjacent segment disease and elimination of the risk of pseudoarthrosis formation. Initial studies of cervical arthroplasty have shown promising results, however, the ideal design characteristics for disc replacement constructs have not been determined. The current study seeks to quantify the differences in the shock absorption characteristics of three commonly used materials in cervical disc arthroplasty.

Methods

Three different nucleus materials, polyurethane (PU), polyethylene (PE) and a titanium-alloy (Ti) were tested in a humidity- and temperature-controlled chamber. Ten of each nucleus type underwent three separate mechanical testing protocols to measure 1) dynamic stiffness, 2) quasi-static stiffness, 3) energy absorption, and 4) energy dissipation. The results were compared using analysis of variance.

Results

PU had the lowest mean dynamic stiffness (435 ± 13 N/mm, P < .0001) and highest energy absorption (19.4 ± 0.1 N/mm, P < .0001) of all three nucleus materials tested. PU was found to have significantly higher energy dissipation (viscous damping ratio 0.017 ± 0,001, P < .0001) than the PE or TI nuclei. PU had the lowest quasi-static stiffness (598 ± 23 N/mm, P < .0001) of the nucleus materials tested. A biphasic response curve was observed for all of the PU nuclei tests.

Conclusions

Polyurethane absorbs and dissipates more energy and is less stiff than either polyethylene or titanium.

Level of Evidence

Basic Science/Biomechanical Study.

Clinical Relevance

This study characterizes important differences in biomechanical properties of materials that are currently being used for different cervical disc prostheses.

keywords: 
Cervical disc arthroplasty, Shock Absorption, degenerative disc disease, Cervical Spine