Table 3

Cage motion results for pedicle/screw rod and anterior-lateral plate specimens with endplate fracture during interbody cage placement

Endplate fracture location and levelLateral cage translation (mm)
F/ELBAR
A-PM-LI-SA-PM-LI-SA-PM-LI-S
Pedicle screw/rod Inferior endplate at L4-L50.50.10.30.40.10.53.31.21.0
*PSR groupavg.(StDev)
n = 17 levels
0.40.10.20.20.10.20.80.30.2
(0.1)(0.1)(0.1)(0.1)(0.1)(0.2)(0.4)(0.3)(0.1)
Lateral plate/screw
  AInferior endplate at L3-L40.10.10.13.20.51.80.50.20.2
  Inferior endplate at L4-L50.10.00.10.20.00.20.40.10.3
  Superior endplate at L4-L50.30.10.70.40.82.12.11.50.5
  BInferior endplate at L3-L40.50.10.00.20.10.00.70.20.2
*ALP groupavg.(StDev)
n = 14 levels
0.30.10.20.20.20.20.80.20.4
(0.2)(0.1)(0.4)(0.1)(0.4)(0.2)(1.0)(0.3)(0.4)
  • Abbreviations: avg., average; StDev, standard deviation. Lateral cage translations were measured during cyclic loading along theanterior-posterior (A-P), medial-lateral (M-L), and inferior-superior (I-S) axes for each loading mode.

  • * Average cage translations for the respective treatments (PSR and ALP) inspecimens with intact endplates.