Table 2

Gravimetric Weight Loss and Wear Debris Particle Sizes (Over 10 Million Cycles of Simulated Loading)

Average Volume Percent of Small (Phagocytosable) Particles
FluidWeight Loss (mg)Type of Debris (EDAX)Average Diameter (Volume)Average Diameter (Number)<1µm<2µm<5µm
Saline/SerumNot detectableCo-Cr alloy48.4µm3.2µm2.6%6.2%17.6%
Saline200Cr-Co-Oxide16.2µm0.2µm17.8%28.7%44.0%