Range of motion comparison between the two wear standards
IS0 18192-110 | ASTM F2324-0511 | |
---|---|---|
Cervical | ||
Flexion extension | ±7.5° | ±7.5° |
Lateral bending | ±6.0° | ±6.0° |
Axial rotation | ±4.0° | ±6.0° |
Lumbar | ||
Flexion extension | ±6°/ − 4° | ±7.5° |
Lateral bending | ±2.0° | ±3.0° |
Axial rotation | ±2.0° | ±6.0° |