Introduction

Pedicular screw placement is a standard procedure for spinal fusion in the lumbar and thoracic spine. In recent years image guidance has increased the accuracy of pedicular screw placement [13]. Recent developments include a spinous process-mounted miniature robot, the SpineAssist™ (Mazor Robotics, Caesarea, Israel). Pechlivanis and colleagues [4] demonstrated the high accuracy of this robotic system. Nevertheless, with regard to the great variability of pedicle screw deviation rates given in literature (1–18% [57]), it remains difficult for the surgeon to estimate the benefit of image guidance without a control group.

Likewise, a number of techniques for percutaneous pedicle screw placement have been introduced, mostly with favorable initial results [812]. However, no clinical study directly comparing the outcome of percutaneous and open midline approaches has yet been published in the English literature.

We here present a single institution retrospective analysis of perioperative results following lumbar and thoracic pedicle screw insertion using three different techniques.

Materials and methods

Patients

112 consecutive patients who underwent pedicular screw placement between January 2006 and August 2009 were analyzed regardless of the indication for spinal fusion.

Surgery

Pedicle screw placement was performed using three alternative techniques: patients received pedicle screws either using conventional 2-D fluoroscopic guidance (BV Endura, Philips, Hamburg, Germany) via a midline incision (termed “conventional”), or using an robotic-guided technique (SpineAssist™ Mazor Robotics, Caesarea, Israel). These robotic-guided procedures were either performed via a midline incision (open robotic-guided), or via a percutaneous approach. For percutaneous surgery separate 2 cm incisions for each screw and a 3–4 cm midline incision, to attach the robot to the spinous process by a clamp, were performed (percutaneous robotic-guided, Fig. 1).

Fig. 1
figure 1

Pictures a percutaneous robotic-guided procedure, as performed in 35 patients in this study. a1 shows the SpineAssist™ robot fixed to the spinous process by a clamp. The robot is already in the final position as planned by the surgeon on the planning-workstation. The arm which will guide drilling and guiding-wire insertion is being fixed to the robot. a2 shows placement of a guiding tube and marking of the skin at the planned entry point for pedicle screw placement. In a3 the skin has been incised at the marked spot and the guiding tube is inserted through the muscles to the entry point in the vertebral bone. a4 shows robotic-guided drilling of a transpedicular burr hole, a5 positioning of the guiding-wire and a6 percutaneous insertion of the screw. b shows a postoperative axial CT scan of the same screw and c displays the surgical wound postoperatively. The central incision was used for exposure of the spinous process and fixation of the robot and for placement of two PLIF cages. The four smaller incisions (two on each side) are the entry points of the percutaneously placed pedicle screws. The bar in the left lower corner of c is a cm scale

In percutaneous cases additional PLIF- and TLIF-cage implantation and decompression of the spinal channel by laminotomy or laminectomy could be performed via the 3–4 cm midline incision when indicated (Fig. 2).

Fig. 2
figure 2

a Shows the surgical scar (left image) and postoperative CT scan (one sagittal and two axial reconstructions) after a robotic-guided percutaneous procedure (instrumentation from L2 to L4 because of osteoporotic L3 fracture). b Shows the same for a similar case (L3–L5 because of osteoporotic L4 fracture) operated by conventional pedicle screw positioning via a midline incision. The bars in the left lower corners of the photographs are cm scales

A pre-operative planning-CT was performed in all cases (16-slice MSCT-scanner, Aquilion™, Toshiba Medical Systems, Neuss, Germany). In robotic-guided cases 3-D reconstructions of this CT were used to plan the screw position using the SpineAssist™ planning-software. Registration of the SpineAssist™ robot was performed by matching the pre-operative CT scan with two intraoperative fluoroscopies (AP- and 60° oblique images; BV Endura, Philips, Hamburg, Germany, see [6] for further details).

Six surgeons performed operations; all of them applied conventional and robotic-guided techniques. In 2006 all patients were operated conventionally, while robotic guidance was introduced in 2007 and implants for percutaneous surgery were introduced 2008. Following the introduction of the new techniques robotic-guided percutaneous surgery was preferred if the implants and instruments were available. The perioperative management, however, was not changed during this period.

Implants

Four types of pedicle screws were used: Tango™ (Ulrich medical, Ulm, Germany) and TRSH 3D™ (Medtronic, Tolochenanz, Switzerland) in conventional procedures, Titan Expertise™ (Peter Brehm, Weisendorf, Germany) in robotic-guided and Icon™ (Blackstone Medical, Laichingen, Germany) in percutaneous procedures. Screw diameter varied from 4.5 to 7.5 mm.

Assessment of screw position

Screw position was assessed on CT scans obtained 1–4 days after surgery. Sagittal and axial reconstructions were analyzed by an investigator blinded to the insertion technique. The assessment was repeated 11 times using a millimeter scale; the relative numbers given in % represent the average values of these measurements. Screws entirely in the bone were graded 0, grade 1 described an encroachment of the cortical bone, grade 2 a deviation of less than 3 mm, grade 3 deviations from 3 to 6 mm and grade 4 for deviations of more than 6 mm (Fig. 3).

Fig. 3
figure 3

Displays the system applied in this study for grading of pedicle screw deviations

Assessment of X-ray exposure

Pre- and postoperative CT scans were performed routinely. The total intraoperative X-ray exposure was recorded and divided by the number of screws placed. For robotic-guided cases this included the registration of the SpineAssist™.

Collection of clinical data

Intraoperative adverse events, operating time, postoperative hospitalization, postoperative administration of opioids, adverse events and reoperations were assessed. Analysis was based on routine-examinations directly postoperative, before discharge and about 12 weeks after surgery. All information was extracted from the patients’ charts and CT scans.

Exclusion from analysis

Patients with spondylodiscitis were excluded from analysis hospitalization and infection rates, because their routine treatment included 3 weeks of i.v. antibiotics in hospital and they had infections prior to surgery.

Patients who were on opioids at hospital admittance were excluded from the analysis of postoperative opioid administration. Patients with incompletely documented data had likewise to be excluded from the concerned analysis (Table 1).

Table 1 The exact number of patients who have been included in the analysis of each parameter

Statistics

Statistical significance was tested by student t test when a Gaussian distribution was expected; in samples for which a normal distribution could not be assumed the Wilcoxon test was used instead. For unquantified data (nominal scale) the χ 2 test was used. Significance level was set α = 0.05.

Results

Average patient age was 63.1 years, gender ratio (m/f) was 52/60. Average BMI was 26.8. These values did not significantly differ between the 3 (sub)groups (Table 2).

Table 2 An overview of patient characteristics for the analyzed groups and subgroups

A total of 536 pedicle screws were placed in 112 patients. 286 screws in 57 patients were implanted conventionally and 250 (55 patients) in robotic-guided procedures. 94 of the robotic-guided procedure were performed via a midline incision (20 patients), while 156 screws (35 patients) underwent percutaneous procedures.

Accuracy of pedicle screw placement

94.5% of robotic-guided and 91.5% of conventionally placed screws were graded 0 or 1. Mild deviations (grade 2) were observed in 4.3% of robotic-guided and 5.0% of conventionally placed screws. Moderate (grade 3) and severe deviations (grade 4) were observed in 0.7 and 0.4% of robotic-guided and 1.9 and 1.6% of conventionally placed screws (Table 3). This was statistically significant regardless of whether grade 0 and grade 1 or exclusively grade 0 was rated as correct placement (Table 4). Results for percutaneous and open robotic-guided procedure did not differ obviously. Percutaneous procedures were significantly more accurate if only grade 0 was considered correct (p values, robotic-guided/conventional 0.00001; percutaneous/conventional 0.001, open robot-guided/convent. 0.00001; percut./convent., 0.001), while open robotic-guided procedures were superior if grade 0 and 1 were analyzed together (Table 3).

Table 3 The relative frequency of deviation grades (0–4, compare Fig. 3) of pedicle screws for the different (sub-)groups (robotic-guided, percutaneous robotic-guided, open robotic-guided and conventional procedures)
Table 4 The relative numbers of deviations for pedicle screws in different regions of the spine (thoracic, lumbar and sacral)

Intraoperative X-ray exposure

During robotic-guided procedures the average X-ray exposure was 34 s compared to 77 s in conventional cases. Average X-ray exposure for percutaneous robotic-guided screws was lower than X-ray exposure during open robotic-guided procedures (27 s compared to 43 s). The differences between robot-assisted (both subgroups together and separately) and conventional procedures were statistically significant, the difference between percutaneous and open robot-assisted procedures was not statistically significant (p values, robotic-guided/convent. 0.0001; percut./convent. 0.001, open robot-guided/convent. 0.023).

Duration of surgery

The duration of surgery documented in the patients’ records always included all surgical steps from positioning to wound closure (including registration of the SpineAssist™, PLIF- or TLIF-cages implantation, laminectomy, etc.). This operation time was divided by the number of screws placed. Average time per screw was found to be 59.1 min in robotic-guided (57.0 min in percutaneous and 65.2 min in open robotic-guided procedures) and 52.9 min in conventional procedures. These differences were not statistically significant.

Postoperative administration of opioid analgesics

Postoperative routine included administration of NSAIDs. Novaminsulfone was added on request. When this regimen failed, opioids were added. 45.45% of the robotic-guided (37.5% of percutaneous and 66.6% of open robotic-guided procedures), and 88.9% of conventionally operated patients who were not on opioids pre-operatively required postoperative administration of opioid analgesics (Table 5. Differences between robot-assisted and conventional and percutaneous robot-assisted and conventional (sub)groups were statistically significant (p = 0.0002 and 0.004).

Table 5 The relative and absolute numbers of patients who required new opioid analgesics after surgery, the rate of adverse events (for intraoperative adverse events, wound healing disorders, postoperative infections and cerebrospinal fluid fistulas) the average duration of postoperative hospitalization and the rate of secondary surgical interventions

Adverse events

Intraoperative adverse events (1 major hemorrhage, 6 dural tears) were observed in 4.7% of robotic-guided cases (6.1% of the percutaneous and 0% of the open robotic-guided cases) and 9.1% of conventional procedures.

Postoperative cerebrospinal fluid fistulas were not observed following robotic-guided procedures (0%), while 6.1% of the conventionally operated patients developed this complication. Postoperative infections occurred in 2.7% of robotic-guided (0% after percutaneous and 10.0% after open robotic-guided procedures), while it was observed in 10.7% of the conventional procedures (Table 5). Statistical analysis showed significance between robot-assisted and conventional procedures for intraoperative adverse events only (p = 0.04).

Postoperative hospitalization

Routinely patients were scheduled to remain ten postoperative days in hospital. Average hospitalization was 10.6 days following robotic-guided fusion (10.1 days for percutaneous, 11.6 days for open robotic-guided procedures) and 14.6 days following conventional procedures (Table 5). Differences between the robot-assisted and conventional and percutaneously and conventional (sub)groups were statistically significant (p values, robot-guided/convent. 0.009; percut./convent. 0.012).

Revision surgery

Revision of misplaced screws was performed in 8 cases; 1.0% robotic guided (0.3% percutaneous and 6.3% open robotic-guided) and 12.2% conventional procedures (Table 5). Secondary sutures (mostly under local anesthesia on the ward) had to be placed because of wound healing disorders/infections in 10 cases (0.6% percutaneous procedures, 12.6% open robotic-guided procedures and 12.2% conventional procedures). 1 case of dislocated PLIF cage had to be revised (open robotic-guided). No statistical significances were found.

Discussion

Perioperative clinical outcome and accuracy of pedicle screw placement in cohorts of robotic-guided (open and percutaneous) and conventionally placed pedicle screws were assessed. We focused on intra- and perioperative outcome, because the advantage of image guidance should show during surgery, respectively, on a postoperative CT scan, while the effect of a minimally invasive approach will be strongest during wound healing. Long-term results (fusion rate, etc.) were not included as it can be expected that these will be similar for all pedicle screws regardless of the surgical approach applied.

Assessment of accuracy of screw positions

Assessment of pedicle screw position in replicate determinations by a single investigator blinded to the insertion technique was performed in order to minimize the effect of investigator-dependent errors. Assessment was based on the evaluation scale proposed by Wiesner [8] and Schizas [3].

Many clinicians accept deviations up to 2 or 3 mm because minor deviations rarely become symptomatic. Lonstein et al. [5] report in their meta-analysis of 4,790 screws of 5.1% screws breaching the cortical bone. No more than 0.2% of these caused neurological symptoms. Nevertheless, from time to time surgeons will be confronted with remaining/new-onset symptoms in the presence of a minor screw deviation and face the dilemma whether to reoperate or not. Therefore, we here recorded all minor and major deviations.

Image guidance

Following the literature conventional screw placement results in a deviation rate of 1–18% [57], while other authors described a significant reduction of this deviation rate when image guidance is applied (2.7–3.9% [1315]).

In percutaneous procedures deviation rates tend to be higher (between 6.6 and 80% [2, 3, 12, 16]) and image guidance has likewise been shown to reduce this [16].

Presently, a great variety of systems offering spinal image guidance are under research, clinical evaluation or already in routine clinical use. Among the proposed techniques are electromagnetic navigation [17], intraoperative 3-D fluoroscopy and CT navigation [16, 18], and percutaneous reference frames [19] and robotic-guided surgery [4].

Accuracy rates

Pechlivanis and colleagues [4] report in their recent study using SpineAssist™ of 91.7% of screws entirely within the bone, while further 6.8% of the screws showed deviations of less than 2 mm. This corresponds to our results (94.6% grade 0 and 1 and 4.3% grade 2).

The accuracy of conventionally placed pedicle screws in our study was 91.4% grade 0 or 1 and further 5.0% grade 2. This corresponds to given in literature [57].

In accordance with the literature [16], in our study the accuracy of image-guided pedicle screw placement was significantly higher than in conventional procedures (about 3.1%). In robotic-guided procedures the surgical approach did have any obvious influence on the accuracy of screw positions. This contrasts with the situation in conventional procedures, as discussed above [2, 3, 16].

X-ray exposure

Intraoperative X-ray exposure was significantly reduced by application of robotic guidance. Pre- and postoperative CT scans were performed routinely, so that the SpineAssist™ planning-CT did not cause additional X-ray exposure. Furthermore, the preoperative planning-CT did not affect the surgical team which, in contrast to the patient, faces the intraoperative radiation frequently.

Interestingly, we observed a decrease of intraoperative X-ray doses during percutaneous robotic-guided procedures compared to open robotic-guided procedures. The best explanation is, however, our learning curve with the SpineAssist™. The first robotic-guided procedures in 2007 were open robotic-guided and the decreased radiation during percutaneous procedures might reflect a growing confidence in the SpineAssist™ in 2008.

Clinical parameters

Duration of postoperative hospitalization, postoperative opioid administration (indicator of severe pain), infection rate and rate of screw revisions were better in the robotic-guided group. However, the percutaneous subgroup did still better, while the open robotic-guided subgroup resembled more the conventional cases. This indicates that the perioperative clinical course depends less on the application of image guidance than on the surgical approach (percutaneous versus open).

Percutaneous pedicle screw placement

In literature many authors report good clinical results following percutaneous pedicle screw placement for various indications (trauma [9], spondylodiscitis [20] and spondylolisthesis [10, 11]). Accordingly, several studies measuring serum creatine kinase levels after surgery indicated less muscular trauma in percutaneous operations [21, 22]. Cadaver and animal studies comparing percutaneous and open midline approaches supported these findings [23, 24].

However, studies directly comparing clinical outcome after percutaneous and open pedicle screw placement in humans are not yet published in the English literature. Grass and colleagues report in a German journal “Der Unfallchirurg” of 57 patients with thoracolumbar fractures. EMG examination proved greater trauma in open procedures and blood loss was significantly higher. Two of the 35 patients operated percutaneously needed a revision. However, the study focuses on intraoperative issues and accuracy of screw placement, further clinical data is not given in this interesting article [9]. Another study, which appeared in the Chinese language journal “Zhonghua Wai Ke Za Zhi” reports of a prospective study in 60 patients with thoracolumbar burst fractures, who had been treated with either percutaneous or open pedicular screw placement [25]. The authors conclude that “pedicle screw fixation for thoracolumbar fracture has the advantage of less trauma, quicker recovery and better esthetic outcomes, however, it has the same results with the traditional open produce after 2 years of surgery”. Our study likewise indicates that clinical outcome following percutaneous procedures might be superior to that following an open midline approach.

Shortcomings of the study

The underlying pathologies were heterogeneous. Posterior stabilization with pedicle screws and rod systems was frequently combined with other surgical procedures. Therefore, the assessed parameters were not exclusively related to pedicle screw implantation, but may partly reflect additional surgical procedures. Some cases had to be excluded from the analysis of one or several parameters because of the underlying disease. This influences the quality of our data and the conclusions derived from it.

Besides this study spans from 2006 to 2009 and the robot-assisted (percutaneous) subgroup represents the most recently operated cohort. Patients were operated by six different surgeons and the data includes all their learning curves with percutaneous approaches, the SpineAssist™ and spinal fusion surgery in general. Additionally, no randomization was performed, so that an entry bias might further confound this study.

Conclusions

Robotic-guided pedicle screw implantation resulted in accurate placement in 94.5% compared to 91.4% in conventional screw placement. Duration of intraoperative radiation was significantly lower in robotic-guided compared to conventional procedures.

Postoperative administration of new opioid analgesics, duration of postoperative hospitalization and rates of adverse events seem to be lower in percutaneous robotic-guided procedures compared to open robotic-guided and conventional procedures.