Pelvic Fixation Configuration Count Reported Complaints,
n (%)Single osseointegrative implant 4381 1 (0.02) 2+ osseointegrative implants 1110 38 (3.40) Osseointegrative implant + additional fusion device 1394 13 (1.35) Trauma fixation 22 - Complaint Count Days to Event,
Mean (SD)Procedural Complaint Rate, % (N = 6907) Set screw disassociation 23 133.0 (102.2) 0.33 Implant loosening 10 375.6 (103.1) 0.15 Infection (unrelated to implant) 8 215.4 (107.7) 0.12 Implant sleeve separation 5 116.8 (163.4) 0.07 Lucency/halos 4 414.0 (222.7) 0.06 Malpositioned implant 2 178.5 (212.8) 0.03 Indication Early Revisions Late Revisions Count Days to Event,
Mean (SD)Count Days to Event,
Mean (SD)Lucency/haloing 0 - 3 487.7 (163.4) Set screw dissociation 8 47.9 (14.3) 11 199.3 (103.1) Implant loosening 0 - 10 375.6 (103.1) Malpositioned (asymptomatic) 1 28.0 (0.0) 1 329.0 (0.0) Infection (unrelated to implant) 1 90.0 (0.0) 6 375.6 (103.1) Total 10 50.1 (19.9) 31 299.5 (140.9)